+852 2866 8995 General Enquiry: cs@medikonia.com | Support: support@medikonia.com

RayBiotech Denosumab ELISA

Bridging ELISA. Catalog #: EAD-Deno

Product description Order information Resources Panel

Denosumab ELISA

This kit is a snadwich-based bridging ELISA for the quantitative measurement of Denosumb in serum, plasma, and cell culture supernatants. 6 - 8 week lead time.

 

Specifications

Size: 1 Plate Kit, 2 Plate Kit, 5 Plate Kit

Targets Detected: Denosumab

Protein Name / Synonyms: Denosumab

Quantitative / Semi-Quantitative: Quantitative 

Compatible Sample Types: Cell Culture Supernatants, Plasma, Serum

Solid Support: 96-well Microplate

Method Of Detection: Colorimetric

Design Principle: Sandwich-based

Research Area: Antibody Drug Development, Drug Antibody (DA) Detection

 

Applications Notes

Kit Components

  • - Pre-Coated 96-well Strip Microplate

  • - Wash Buffer

  • - Lyophilized Standard

  • - Biotinylated Detection Antibody

  • - Streptavidin-Conjugated HRP

  • - Stop Solution

  • - Assay Diluent

 

Other Materials Required

  • - Microplate reader capable of measuring absorbance at 450 nm

  • - Precision pipettes to deliver 2 µl to 1 ml volumes

  • - Adjustable 1 - 25 ml pipettes for reagent preparation

  • - 100 ml and 1 liter graduated cylinders

  • - Absorbent paper

  • - Distilled or deionized water

  • - Log-log graph paper or computer and software for ELISA data analysis

  • - Horizontal orbital microplate shaker

  • - Tubes to prepare standard and sample dilutions

 

Protocol Outline

  • 1. Prepare all reagents, samples and standards as instructed in the manual.

  • 2. Add 100 µl of standard or sample to each well. Incubate 2 hours at room temperature.

  • 3. Add 100 µl prepared biotinylated antibody to each well. Incubate 1 hour at room temperature.

  • 4. Add 100 µl prepared Streptavidin solution. Incubate 45 minutes at room temperature.

  • 5. Add 100 µl TMB One-Step Substrate Reagent to reach well. Incubate 30 minutes at room temperature.

  • 6. Add 50 µl of Stop Soution to each well. Read at 450 nm immediately.

 

Storage/Stability

The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. The kit may be stored at 4°C for up to 6 months. For extended storage, it is recommended to store at -80°C.

Catalog #: EAD-Deno-1 -- 1 Plate Kit

Catalog #: EAD-Deno-2 -- 2 Plate Kit

Catalog #: EAD-Deno-5 -- 5 Plate Kit
 

Please email us at cs@medikonia.com for any enquiry. To place an order, please include the catalog number(s) of the product(s) in the email.